KAJIAN AWAL PENGEMBANGAN METODE ANALISIS ION SULFAT BERBASIS ONE DROP FLOW INJECTION ANALYSIS (OD-FIA)

The analytical methods are generally high-cost, time and chemicals consuming. The usage of these chemicals makes the laboratory becomes the source of environmental pollutions. Flow Injection Analysis (FIA) method is a solution for reducing the high usage of chemicals since its measurement is less...

全面介紹

Saved in:
書目詳細資料
主要作者: Riptono, Agus
格式: Final Project
語言:Indonesia
主題:
在線閱讀:https://digilib.itb.ac.id/gdl/view/71713
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
機構: Institut Teknologi Bandung
語言: Indonesia
實物特徵
總結:The analytical methods are generally high-cost, time and chemicals consuming. The usage of these chemicals makes the laboratory becomes the source of environmental pollutions. Flow Injection Analysis (FIA) method is a solution for reducing the high usage of chemicals since its measurement is less time consuming and it does not need to reach the real system equilibrium. An analysis protocol for sulfate ion using one drop FIA method with spectrophotometric detection technique using optical fiber equipment was developed and evaluated. The evaluation results in a high analytical performance of the method, shown by the reproducibility values of 7.14% and 1.96% for concentration of 25 ppm and 125 ppm, respectively. Calibration curve showed a good linearity at the concentration range of 25 to 125 ppm with the approximate correlation value of 1. This method was applied to analyze the sulfate ion concentration in water from Jatiluhur dam.